Thursday, July 25, 2024

Testing another Aliexpress Antenna

Below the results of the measurement of another Antenna

I did buy this antenna some time ago on AliExpress.


Wednesday, July 24, 2024

Testing Quansheng Antenna

When testing the gate fence antenna (see a previous post) i also wanted to compare the measured result with some other antenna's. This is the result of a quick measurement of an antenna that came with of one of my Quansheng 2m/70cm radio's.



Tuesday, July 23, 2024

Gate fence electronics dismantling -2- Testing gate fence antenna

Testing a gate fence antenna with a vector network analyzer. This is a picture of the antenna of the original receiver for the fence controller after some cleaning. ( See previous blogpost)

As i have a vector network analyzer (vna) i wanted to test the antenna to check on what frequencies it can be (re)used. I soldered an sma connector to the coax wire, calibrated the vna and did a quick scan over a relative broad range.

See below the results displaying Smith chart, Quality factor and swr (VSWR). The lower the swr the better. The minima in the swr curve give an indication of optimal frequencies.

I also did the same quick scan on some other antenna's and post the blog reports with the results.  Therefore this blogpost is in fact part of two different series of blog posts.

-1- Dismantling the electronics of a gate fence. ( #GateFence )
-2- Some antenna's quick tested ( with a vector network analyzer (vna) ).


Monday, July 22, 2024

Gate fence electronics dismantling -1- Main control and rx board

In this post some pictures of electronics in a 220 Volt gate controller to control 220 Volt motors.After many years of service this device was not working reliable anymore and i needed to replace it. This blogpost is also to document the old situation.

Top left you see a transformer top right are battery holders (not used) . In the middle of the picture you see the main controller board. Also you  can see a part of the antenna. However this remote control was not used.

The cable bundle on the right form a separate circuit with cables going to a light sensor for a lamp.

In this picture the mains to the control board is disconnected .(The wires where temporary placed in additional screw terminals connected to nothing to put the fence control temporary out of service).

Original the connection was to the left two connections where also the white wire on the left is connected. The white wire is to power a 433 MHz remote control switch box (bottom left). From this control box go wires to contacts on the main controller board to external contact to open and close the gate. Below a closeup of the control board after disassembling.

On the right is the connector for the receive module. Below some pictures of this module.


As already mentioned the control board is completely replaced with a new (different) board. I also removed the antenna and tested it with a vector network analyzer. You can see the bottom part of the antenna in the first picture of this post top right.
Results of antenna measurements will come in a next blogpost.

Saturday, July 6, 2024

Measuring Resistance of Fuses with a Milliohm Meter and Kelvin Clips -4- (Self Resettable fuse )

Recently i received this 30A self resettable fuse via AliExpress.


As you can see on the pictures on one side is printed:

A                        (i don't know yet what this means)
Suppl. Prot. 
LINE  LOAD (to mark te connections).

On the other side of the self resettable fuse has the tekst

ZHONGZUI 78 Series 30A
125/250 VAC 50/60 Hz  50VDC 
TC3OL1/U3

And symbols suggesting that the fuse had passed some tests.

And yes i tested it with my Milliohm meter by connecting the Kelvin Clips.

The measured resistance was only 0.002 Ohm

I don't know what the resistance will be under normal use contions, however is was happy surprised with this low resistance !! 

Wednesday, June 12, 2024

Measuring Resistance of Fuses with a Milliohm Meter and Kelvin Clips -3- (More 12 Volt Fuses)

I received more 12 fuses and in this this blogpost i am going to present the results of the 12 Volt fuses. My intention was to publish after my post about the 250V glass fuses some evaluation of the data. However i received from Aliexpress a nice Anderson Powerpole connection block "DC 36V AP-8S 8 Channel Power Distributor For HF Radio Communication Power Supply Splitter Anderson Powerpole Screw Fixing" with more 12 Volt fuses. 

As i only had a few results of 12V fuses in my first post about the car fuses and i was curious about these Chinese fuses i gave more priority to measure the 12V fuses.

I want to do a separate blogpost about a review of this Anderson Powerpole block (and it also will appear in a mailbag blog) as i think it is a real nice power block that i bought for a decent price.  Not only does did it had a set of fuses in the block, also a spare set of fuses was included. 
So i now additional have 2 x ( 1x 40A + 3x 25A + 2x 15A + 1x 10A + 1x 5A ) fuses.

In the table below are also the results from the set car fuses already mentioned in the previous blog about the 12 V car fuses (= S1 [set1] ) and the fuse that came with the single fuse holder (= H1)

The measurement with this single fuse holder are not included as i wanted to focus this post on the fuse not on the fuse holder(s). 

Resettable Fuse Circuit Breaker

Via Aliexpress i also got a 12V 24V DC Car Truck Audio Resettable Fuse Circuit Breaker 30A Circuit Breaker. (see picture below) 

Resettable Fuse Circuit Breaker

A nice thing about this fuse is that you can reset it and it can be used as circuit breaker. ('Resettable Fuse Circuit Breaker'). I was curious about this fuse so i ordered one and included the result in the table below.  Pressing the circuit breaker and resetting the fuse (several times) gave each time the same resistance.
Spoiler alert: Relative to the other 30A fuse the resistance seems higher. However the resistance is still low.

Safety

Be aware that these resetable fuse circuit breakers (and also the normal 12V car fuses from AliExpress) have a bad reputation and can be unsafe.

The Carpoint fuses that i did buy at the Gamma  in The Netherlands are also made in China. (See packaging bottom right). 

Measurement

I removed fuses (one by one) from the Anderson Powerpole block and measured the resistances ( Ab to Ab7 [the number corresponds with the number on the Anderson Powerpole block ] ). I was a little difficult to remove these fuses. 

The spare fuses that came with the Anderson Powerpole block ( Ax to Ax7 ) where also measured.

The results

Table 1: Results 12V fuses (Including previous results, sorted on rated current). 
Fuse Color Rated CurrentMeasured
Resistance
Calculated
Voltage drop
at 100 mA
Calculated
Voltage drop
at rated current
Calculated Voltage drop
at rated current in %
relative to 12 Volt
Ab7 Biege 5 0.014 0.0014 0.070 0.58%
Ax7 Biege 5 0.013 0.0013 0.065 0.54%
S1 Brown 7.5 0.009 0.0009 0.068 0.56%
Ab6 Red 10 0.006 0.0006 0.060 0.50%
Ax6 Red 10 0.006 0.0006 0.060 0.50%
H1 Red 10 0.007 0.0007 0.070 0.58%
S1 Red 10 0.006 0.0006 0.060 0.50%
Ab4 Blue 15 0.004 0.0004 0.060 0.50%
Ab5 Blue 15 0.004 0.0004 0.060 0.50%
Ax4 Blue 15 0.005 0.0005 0.075 0.63%
Ax5 Blue 15 0.005 0.0005 0.075 0.63%
S1 Blue 15 0.004 0.0004 0.060 0.50%
S1 Yellow 20 0.003 0.0003 0.060 0.50%
Ab1 Clear 25 0.002 0.0002 0.050 0.42%
Ab2 Clear 25 0.003 0.0003 0.075 0.63%
Ab3 Clear 25 0.003 0.0003 0.075 0.63%
Ax1 Clear 25 0.003 0.0003 0.075 0.63%
Ax2 Clear 25 0.003 0.0003 0.075 0.63%
Ax3 Clear 25 0.003 0.0003 0.075 0.63%
S1 Clear 25 0.002 0.0002 0.050 0.42%
Resettable
 Fuse Circuit Breaker
 (*1) 30 0.004 0.0004 0.120 1.00%
S1 Green 30 0.002 0.0002 0.060 0.50%
Ab Amber 40 0.001 0.0001 0.040 0.33%
Ax Amber 40 0.001 0.0001 0.040 0.33% 

Summary

Higher rated fuses seem to have a lower resistance.

Between 12 Volt fuses of the same rating is not much variation in the resistance.

The 30 A Resettable Fuse Circuit Breaker seems to have a higher resistance (and higher voltage drop) than a standard 30A 12V fuse.

Also, compared to the 250 Volt fuses in my previous post these 12V car fuses have a much lower resistance !!!

Tuesday, June 11, 2024

Measuring Resistance of Fuses with a Milliohm Meter and Kelvin Clips -2- (250V Fuses measurement results )

Ensuring the integrity and reliability of 250V fuses is crucial for maintaining the safety and functionality of high-voltage electrical systems. In this post, I'll share my experience measuring the resistance of 250V fuses using an affordable milliohm meter equipped with Kelvin clips, highlighting the key considerations and benefits of this method.

Why Measure Fuse Resistance?

Regularly measuring the resistance of 250V fuses is essential for several reasons:

1. Ensuring Reliability:

Over time, fuses can develop higher resistance due to aging or exposure to harsh conditions.

2. Troubleshooting:

Identifying a fuse with abnormal resistance can help diagnose and prevent circuit issues.

3. Preventive Maintenance: 

Regular measurements allow for the replacement of fuses before they fail, avoiding potential downtime or damage.

Tools Needed

For this task, I used:

- Milliohm Meter:

 Capable of measuring low resistance values accurately.

- Kelvin Clips: 

Four-wire probes that eliminate the influence of lead and contact resistance, providing precise measurements.

The Measurement Process

Here's a quick overview of the process:

1. Calibration:

Ensure the milliohm meter is calibrated using the shorting clip and zeroing function.

2. Connecting Kelvin Clips: 

Attach the clips to the fuse, ensuring a solid, clean connection at opposite ends.

3. Taking the Measurement: 

Power on the meter and read the displayed resistance value.

Voltage Drop Across the Fuse

For a milliohm meter with a measurement current of 100 mA, the voltage drop across the fuse can be calculated using Ohm's Law (V = I × R).

Example Calculation:

- For a measured resistance of 0.005 ohms: 

  - Voltage Drop: V = 100 mA × 0.005 ohms = 0.5 mV

- For a measured resistance of 0.002 ohms: 

  - Voltage Drop: V = 100 mA × 0.002 ohms = 0.2 mV

Change in Resistance Due to Heating

When a fuse is subjected to higher currents, it can heat up, causing its resistance to change:

- Initial Resistance:

The resistance at room temperature is relatively low.

- Heating Effect:

Higher current flow increases temperature, raising the resistance.

- Increased Resistance:

This change can be significant, especially for currents near the fuse’s rated limit.

Measured resistance

I having a long time a box with slow 250V glass fuses with values 50 mA, 160 mA, 200 mA, 315 mA, 500 mA, 800 mA, 1 A, 2 A, 2.5 A. It would be nice to use these fuses as a test-set. 

My box with fuses (open)

I first did think it was needed to measure each fuse several times. However during my first (not documented) test i discovered that remeasuring a fuse id almost gave no variation in the result. (max 1 or 2 in the last digit).  The variation between different fuses of the same rating was much bigger. Therefore i decided to measure all the fuses once and put the results in a table (1).

However there were some limitations: 

- According to the manual, the resistance is measured with a testing current of about 100 mA. ( i did not check this yet.) Therefore i decided to skip to 50 mA fuses as they would be blown an not measured.

- The fuses are relative old. I expect at least 25 years. The nice thing is that it is a set of in total more than 140 fuses of different values and the aging effect of the fuses is included.

- The box with fuses had at (least) two sources. A box with fuses that i did buy many years ago and a similar box with fuses that i received some year ago from someone else. I did merge the content of these boxes a year ago to save some space.

- I noticed some fuses had a resistance far above 1 ohm. (And i expect some of these fuses where bad. I measured with a multi-meter around 200 Ohm and some above 10 Mega Ohm and even some fuses that made no connection at all. I removed these fuses from the box with fuses, kept them apart and did not include the results. I expect the fuses that made no connection at all are blown fuses and i did throw them away.  

- All my 2.5 A fuses had corrosion and the measured resistances where above 1 Ohm. I don't know if this is normal or this was due to aging and/or the corrosion. The strange thing is that all the 2.5 A fuses that had visible corrosion and almost none of the other fuses.  I don't know the source of the corrosion. Perhaps the fuses are of a different material or there was humidity that mainly reached the 2.5 A fuses and not the other fuses. Therefore i excluded all of the 2.5 A fuses.

Results

Below a table with the measured resistance , the calculated voltage drop at 100mA and the voltage drop at the rated current.

Table 1: Measured resistances with calculated voltage drops

Result /
 measurement
Number
 mA Measured
 resistance 
Calculated
Voltage drop
at 100mA
Calculated
Voltage drop
at rated current
1 500 0.447 0.04470 2.235
2 500 0.477 0.04770 2.385
3 500 0.446 0.04460 2.230
4 500 0.434 0.04340 2.170
5 500 0.520 0.05200 2.600
6 500 0.461 0.04610 2.305
7 500 0.510 0.05100 2.550
8 500 0.536 0.05360 2.680
9 500 0.462 0.04620 2.310
10 500 0.432 0.04320 2.160
11 500 0.502 0.05020 2.510
12 500 0.482 0.04820 2.410
13 500 0.452 0.04520 2.260
14 500 0.465 0.04650 2.325
15 500 0.424 0.04240 2.120
16 500 0.456 0.04560 2.280
17 500 0.469 0.04690 2.345
18 500 0.461 0.04610 2.305
19 160 0.579 0.05790 0.926
20 160 0.570 0.05700 0.912
21 160 0.537 0.05370 0.859
22 160 0.640 0.06400 1.024
23 160 0.409 0.04090 0.654
24 160 0.658 0.06580 1.053
25 160 0.664 0.06640 1.062
26 160 0.531 0.05310 0.850
27 160 0.613 0.06130 0.981
28 160 0.568 0.05680 0.909
29 160 0.677 0.06770 1.083
30 160 0.581 0.05810 0.930
31 160 0.604 0.06040 0.966
32 160 0.619 0.06190 0.990
33 160 0.494 0.04940 0.790
34 160 0.615 0.06150 0.984
35 160 0.689 0.06890 1.102
36 160 0.557 0.05570 0.891
37 160 0.409 0.04090 0.654
38 160 2.900 0.29000 4.640
39 1000 0.088 0.00880 0.880
40 1000 0.089 0.00890 0.890
41 1000 0.092 0.00920 0.920
42 1000 0.087 0.00870 0.870
43 1000 0.089 0.00890 0.890
44 1000 0.089 0.00890 0.890
45 1000 0.087 0.00870 0.870
46 1000 0.089 0.00890 0.890
47 1000 0.087 0.00870 0.870
48 1000 0.094 0.00940 0.940
49 1000 0.091 0.00910 0.910
50 1000 0.096 0.00960 0.960
51 1000 0.089 0.00890 0.890
52 1000 0.087 0.00870 0.870
53 1000 0.089 0.00890 0.890
54 1000 0.092 0.00920 0.920
55 1000 0.091 0.00910 0.910
56 1000 0.090 0.00900 0.900
57 1000 0.090 0.00900 0.900
58 315 0.700 0.07000 2.205
59 315 0.562 0.05620 1.770
60 315 0.761 0.07610 2.397
61 315 0.706 0.07060 2.224
62 315 0.852 0.08520 2.684
63 315 0.805 0.08050 2.536
64 315 0.607 0.06070 1.912
65 315 0.723 0.07230 2.277
66 315 0.681 0.06810 2.145
67 315 0.612 0.06120 1.928
68 315 0.646 0.06460 2.035
69 315 0.884 0.08840 2.785
70 315 0.760 0.07600 2.394
71 315 0.719 0.07190 2.265
72 315 0.871 0.08710 2.744
73 315 0.766 0.07660 2.413
74 315 0.681 0.06810 2.145
75 315 0.620 0.06200 1.953
76 315 0.759 0.07590 2.391
77 315 0.602 0.06020 1.896
78 2000 0.039 0.00390 0.780
79 2000 0.039 0.00390 0.780
80 2000 0.039 0.00390 0.780
81 2000 0.042 0.00420 0.840
82 2000 0.038 0.00380 0.760
83 2000 0.037 0.00370 0.740
84 2000 0.035 0.00350 0.700
85 2000 0.703 0.07030 14.060
86 200 0.331 0.03310 0.662
87 200 0.485 0.04850 0.970
88 200 0.545 0.05450 1.090
89 200 0.405 0.04050 0.810
90 200 0.486 0.04860 0.972
91 200 0.584 0.05840 1.168
92 200 0.490 0.04900 0.980
93 200 0.573 0.05730 1.146
94 200 0.472 0.04720 0.944
95 200 0.442 0.04420 0.884
96 200 0.446 0.04460 0.892
97 200 0.508 0.05080 1.016
98 200 0.515 0.05150 1.030
99 200 0.429 0.04290 0.858
100 200 0.489 0.04890 0.978
101 200 0.395 0.03950 0.790
102 200 0.491 0.04910 0.982
103 250 0.268 0.02680 0.670
104 250 0.517 0.05170 1.293
105 250 0.382 0.03820 0.955
106 250 0.388 0.03880 0.970
107 250 0.385 0.03850 0.963
108 250 0.340 0.03400 0.850
109 250 0.348 0.03480 0.870
110 250 0.366 0.03660 0.915
111 250 0.249 0.02490 0.623
112 250 0.322 0.03220 0.805
113 250 0.338 0.03380 0.845
114 250 0.405 0.04050 1.013
115 250 0.263 0.02630 0.658
116 250 0.361 0.03610 0.903
117 250 0.350 0.03500 0.875
118 250 0.247 0.02470 0.618
119 250 0.349 0.03490 0.873
120 250 0.322 0.03220 0.805
121 250 0.342 0.03420 0.855
122 250 0.371 0.03710 0.928
123 250 0.292 0.02920 0.730
124 800 0.151 0.01510 1.208
125 800 0.136 0.01360 1.088
126 800 0.139 0.01390 1.112
127 800 0.127 0.01270 1.016
128 800 0.133 0.01330 1.064
129 800 0.131 0.01310 1.048
130 800 0.141 0.01410 1.128
131 800 0.128 0.01280 1.024
132 800 0.128 0.01280 1.024
133 800 0.133 0.01330 1.064
134 800 0.136 0.01360 1.088
135 800 0.143 0.01430 1.144
136 800 0.127 0.01270 1.016
137 800 0.128 0.01280 1.024
138 800 0.135 0.01350 1.080
139 800 0.136 0.01360 1.088
140 800 0.132 0.01320 1.056
141 800 0.147 0.01470 1.176
142 800 0.132 0.01320 1.056

Some of the fuses gave results that made me doubt if the fuse had indeed the current rating i expected. However i did check the rating it on the fuse! 

Table 2: Summary of measured resistances  (Some outliers are removed.)
Fuse rating (mA) Average resistance  Min resistance  Max resistance
160 0.5797 0.409 0.689
200 0.4756 0.331 0.584
250 0.3431 0.247 0.517
315 0.7159 0.562 0.884
500 0.4687 0.424 0.536
800 0.1349 0.127 0.151
1000 0.0898 0.087 0.096
2000 0.0384 0.035 0.042
 
Fuses with a higher rating seem to have a lower resistance, however even in this box with fuses not always!  I decided to already publish the measured results with a quick summary. In an upcoming blog-post i will do a more in depth evaluation of the results.
My box with fuses (closed)

Conclusion

Measuring the resistance of 250V fuses with a milliohm meter and Kelvin clips is a straightforward yet powerful method to ensure the health of your high-voltage circuit protection devices. Regular checks can help you catch potential issues early, maintaining the integrity and safety of your electrical systems. It was also a good exercise to measure these fuses to check the health and variation between these fuses. This affordable and accurate setup is a valuable addition to any toolkit, providing peace of mind and reliable performance.

Stay tuned for more insights and results from my ongoing experiments and measurements and a more in-depth evaluation of these measured results. Happy measuring!

Monday, June 10, 2024

Measuring Resistance of Fuses with a Milliohm Meter and Kelvin Clips -1- ( Car fuses )

In the world of electronics, ensuring the integrity of fuses is crucial for the safety and reliability of circuits. Fuses act as protective devices, preventing excessive current from damaging components. Over time, however, fuses can degrade, leading to increased resistance, which may compromise their effectiveness. Today, I'll share my experience measuring the resistance of some 12 Volt car fuses using an affordable milli-ohm meter equipped with Kelvin clips. This method provides accurate results, essential for maintaining optimal circuit performance.

Why Measure Fuse Resistance?

Measuring the resistance of fuses serves several purposes:

1. Ensuring Reliability: 

Over time, fuses can develop higher resistance due to aging or exposure to harsh conditions. Regular checks ensure they are still effective.

2. Troubleshooting: 

Identifying a fuse with abnormal resistance can help diagnose and prevent circuit issues.

3. Preventive Maintenance:

Regular measurements allow for the replacement of fuses before they fail, avoiding potential downtime or damage.

Equipment Needed

For this task, I used the following equipment:

- Milliohm Meter: 

A device capable of measuring low resistance values accurately. I used a YMC01 2R handheld Milliohm Meter. For details see my previous blogpost. According to the manual this meter uses a testing current of about 100 mA (i did not check this),

- Kelvin Clips:

 Four-wire probes that eliminate the influence of lead and contact resistance, providing precise measurements. These where included with my meter.

- Fuses:

 For the first test i used some some new 12 Volt Car fuses.

Also i tested a fuse in a fuse holder.
Below a table of the color codes of the car fuses.

The Measurement Process

Here’s a step-by-step guide to measuring fuse resistance with a milliohm meter and Kelvin clips:

1. Calibration

Before starting, it’s essential to calibrate the milliohm meter. Follow these steps:

- Turn on the meter and allow it to stabilize.

- Use the provided shorting clip to connect the Kelvin clips together.

- Zero the meter according to the manufacturer’s instructions. This step ensures that the meter reads zero when there’s no resistance. My cheap meter does not have a zeroing function. H

2. Connecting Kelvin Clips

Attach the Kelvin clips to the fuse. Ensure a solid, clean connection:

- Place the clips at opposite ends of the fuse to measure the resistance accurately.

- Make sure there is no dirt or oxidation on the fuse terminals that could affect the reading.

3. Taking the Measurement

With the Kelvin clips properly connected, take the reading:

- Power on the meter.

- Read the displayed resistance value. The milliohm meter should provide a precise measurement free from lead resistance thanks to the Kelvin clips.

Voltage Drop Across the Fuse

For your milliohm meter with a 0-1 ohm range and a measurement current of 100 mA, you can calculate the voltage drop across the fuse using Ohm's Law (U = I × R).

Example Calculation:

1. Voltage drop at 100 mA : 

   - 100 mA = 0.1 Ampere

   - Measured Resistance: 0.004 ohms

   - Voltage Drop: U = 0.1 A × 0.004 ohms = 0.0004 volts (0.4 mV)

2. Voltage drop at current rate:

   - Fuse rating (e.g. 15 Ampere )

   - Measured Resistance: 0.004 ohms

   - Voltage Drop: U = 15 A × 0.004 ohms = 0.060 volts (30 mV)

2. Relative Voltage drop at current rate:

A voltage drop of 0.060 volts is at 12 Volt the relative voltage drop of 100% * 0.060/12 = 0.50 % 

Change in Resistance Due to Heating

When a fuse is subjected to higher currents, it can get hot, causing its resistance to change. This phenomenon is due to the temperature coefficient of resistance, which describes how the resistance of a material changes with temperature. For most metals, resistance increases as temperature increases.

Key Points:

1. Initial Resistance:

 The resistance of a fuse at room temperature is relatively low.

2. Heating Effect: 

As current flows through the fuse, it heats up. The amount of heat generated is proportional to the square of the current (P = I²R).

3. Increased Resistance: 

As the temperature of the fuse rises, so does its resistance. This change can be significant, especially for currents near the fuse’s rated limit.

Example:

If a fuse rated for 1A is subjected to a current close to its limit, it will heat up. The resistance can increase noticeably, potentially doubling or tripling depending on the fuse material and the current applied.

My Measurement Results

For this first test a set of 12 Volt Car fuses (see pictures above) and a 10A fuse with a fuse holder was used.

Table with Results:

Fuse Rated Current Measured ResistanceCalculated
Voltage Drop (100 mA) 
Calculated
Voltage Drop at rated current
Calculated Voltage drop at rated current in % relative to 12 Volt
10A (without fuse holder)10   A 0.0080.00080.080.67 %
10A fuse in fuse holder10   A 0.0070.00070.070.58 %
Brown 7.5 A 0.0090.00090.06750.56 %
Red 10   A 0.0060.00060.060.50 %
Blue15   A 0.0040.00040.060.50 %
Yellow20   A 0.0030.00030.060.50 %
Clear 25   A 0.0020.00020.050.42 %
Green 30   A 0.0020.00020.060.50 %

Interpreting the Results

- Expected Values:

 Typically, a good fuse should have a very low resistance, often just a few milliohms.

- Anomalies: 

If a fuse shows significantly higher resistance than expected, it might be compromised and should be replaced.

Summary

These resistances are measured with the Milliohm meter. I expect all with around the same current of 100 mA. A higher current will heat up the fuse wire and change the resistance and to get real good conclusions other measurements will be needed. 

The measured resistances of these car fuses where very low (below 0.01 milliohm).

I only had two red 10A fuses (and for the other fuses only one) I measured a resistance of 0.0008 and 0.0006 for these fuses (and 0.0007 ohm for the first red fuse using the fuse holder),

The fuse holder did not add a significant resistance.

The lower rated fuses seem to have a slight higher resistance.

Measuring fuse resistance with a milliohm meter and Kelvin clips is a straightforward yet powerful method to ensure the health of your circuit protection devices. Regular checks can help you catch potential issues early, ensuring your electronics remain safe and functional. By investing in this affordable yet accurate setup, you can maintain the integrity of your circuits and prevent unexpected failures.

Stay tuned for more insights and results from my ongoing experiments and measurements. Happy measuring!